Operation
Alarms Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-81
The Average SHARP screen displays the current average SHARP
concentration and sets the minimum and maximum alarm limits.
Acceptable alarm limits range from 0 to 10000 μg/m
3
or 0 to 10 mg/m
3
. If
the average SHARP concentration goes beyond either the minimum or
maximum limit, an alarm is activated and the alarm (bell) icon appears in
the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms > Conc Alarms >
Avg SHARP
.
AVERAGE SHARP CONC:
ACTUAL 0.0
>MIN 0.0
MAX 10000
RANGE AVG DIAGS ALARM
The Minimum Average SHARP Concentration alarm limit screen is used
to change the minimum average SHARP concentration alarm limit. The
minimum and maximum average SHARP concentration screens function
the same way.
●
In the Main Menu, choose Alarms > Conc Alarms > Avg SHARP >
Min
or
Max
.
AVERAGE SHARP CONC:
ACT MIN: 0.0
SET MIN:
xxxxx
10
0
?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Average Nephelometer screen displays the current average
Nephelometer concentration and sets the minimum and maximum alarm
limits. Acceptable alarm limits range from 0 to 10000 μg/m
3
or 0 to 10
mg/m
3
. If the average nephelometer concentration goes beyond either the
minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms > Conc Alarms >
Avg Neph
.
Average SHARP
Min and Max Average
SHARP Concentration
Limits
Average Nephelometer
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......