Preventive Maintenance
External Pump Exhaust Filter
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
5-9
1.
Turn the Heater OFF: From the Main Menu, choose Instrument
Controls > Set Heater >
Control
. Use
until the value
reads OFF, and press
.
2.
Disconnect RH/temperature connector.
3.
From the SHARP MIB Board J1, remove the five socket-heads from
the top end cap and lift off.
4.
Remove sealing gasket (note: may be stuck to bottomside of top-end
plate).
5.
Carefully vacuum out optic chamber.
6.
Re-assemble in reverse order—note careful placement of the sealing
gasket.
Figure 5–6.
Replacing the External Pump Exhaust Filter
HEPA Filter