Operation
Volumetric Conditions
3-48
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
CHOOSE ACTION:
>NONE
AOUTS TO ZERO <--
AOUTS TO FS
FT COUNT
FT CNTRL
RANGE AVG DIAGS ALARM
PUMP CNTRL
TEMP CNTRL
PRES CNTRL
HT CNTRL A
RH/TEMP CNTRL
ALARM 1
ALARM 2
ALARM 3
The Analog Output Configuration menu displays a list of the analog
output channels available for configuration. Channel choices include all
voltage channels, all current channels, individual voltage channels 1–6, and
individual current channels 1–6 (if the I/O expansion board option is
installed).
●
In the Main Menu, choose Instrument Controls > I/O Configuration
>
Analog Output Confi
g
.
OUTPUT CHANNELS:
>ALLOW OVER/UNDER RANGE
ALL VOLTAGE CHANNELS
ALL CURRENT CHANNELS
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
RANGE AVG DIAGS ALARM
The Allow Over/Under Range screen is used to select whether or not the
analog outputs are allowed to exceed the maximum selected value of 100
mV, 1 V, 5 V, 10 V, or 20 mA or the minimum selected value of 0 V, 0
mA, or 4 mA. By default, this parameter is set to on and 5% over and
under range is allowed for all analog output channels.
●
In the Main Menu, choose Instrument Controls > I/O Configuration
> Analog Output Config >
Allow Over/Under Range
.
Analog Output Configuration
(Select Channel)
Allow Over/Under Range
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......