Installation
Acceptance Testing and Startup Procedures
2-8
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
If the RH sensors are slightly out of tolerance, ±5% RH, the acceptance test
should be classified as marginal. If the sensor performance is outside this
range, please perform a RH-sensor calibration.
For more information about the RH sensor calibration, see the
“
Calibration
” chapter.
It is recommended that the NIST-traceable hygrometer should also
compare well with the RH-measurement used within a gravimetric
laboratory that is part of a compliance program.
Record the Model 5030
i
barometric pressure sensor reading. This value is
in units of mmHg. If necessary, using the pressure unit conversion (
Table
2–1
) to convert your NIST-traceable measurement to units of mmHg for
an appropriate comparison. The Model 5030
i
barometric sensor should
compare within ±10.0 mmHg tolerance of your NIST-traceable barometer.
If the Model 5030
i
sensor performance is within this tolerance, the
acceptance test has passed.
If the sensor is slightly out of tolerance, ±12 mmHg, the acceptance test
should be classified as marginal. If the sensor performance is outside this
range, please perform a sensor calibration.
For more information about the barometric pressure sensor calibration, see
the “
Calibration
” chapter.
The flow rate of the Model 5030
i
should be verified during this acceptance
test. Prior to this test it is important for the previous temperature and
pressure acceptance tests to be completed. Should the ambient temperature,
flow temperature or barametric sensors require calibration, this should be
done prior to the flow rate verification.
●
Press
to return to the Diagnostics menu, and choose
Flows
.
FLOW: 16.67 LPM
RANGE AVG DIAGS ALARM
The Flow screen (read only) displays the volumetric flow rate in liters per
minute. The flow is measured by the internal pressure board sensors.
One-Point Barometric
Pressure Verification
One-Point Volumetric
Flow Rate Verification
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......