Operation
Volumetric Conditions
3-56
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
TABLE POINT 01 USER VAL:
CURRENTLY: 0.00
SET TO: 00000.0
0
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Screen Contrast screen is used to change the contrast of the display.
Intensities between 0 and 100% in increments of 5 are available. Changing
the screen contrast may be necessary if the instrument is operated at
extreme temperatures.
Note
The optimal contrast will change with changes in temperature.
▲
Note
The optimal contrast will change from one LCD to another. If the
LCD is replaced, the contrast may need to be reset.
▲
Note
If the display contrast is not optimal, but the content on the screen is
visible, select Instrument Controls >
Screen Contrast
and adjust the screen
contrast. If the content on the screen is not visible, use the “set contrast 10”
C-Link command to set the screen contrast to mid-range, then optimize
the contrast. See “contrast levels” in Appendix B, “
C-Link Protocol
Commands
” for more information on this command.
▲
●
In the Main Menu, choose Instrument Controls >
Screen Contrast
.
SCREEN CONTRAST:
CURRENTLY: 50 %
SET TO: 55 % ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Service Mode screen is used to turn the service mode on or off.
Turning service mode “ON” locks out any remote actions and allows access
to parameters and functions that are useful when making adjustments or
diagnosing the Model 5030
i
. The service (wrench) icon on the status bar is
shown when service mode is on. For more information about the service
mode, see “
Service Menu
” later in this chapter.
Screen Contrast
Service Mode
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......