C-Link Protocol Commands
Record Layout Definition
B-66
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Send:
set relay open 1
Receive:
set relay open 1 ok
The erec, lrec, and srec layouts contain the following:
●
A format specifier for parsing ASCII responses
●
A format specifier for parsing binary responses
In addition to these, the erec layout contains:
●
A format specifier for producing the front panel display screens
Values are read using either the ASCII or binary format specifiers and
converted to uniform internal representations (32-bit floats or 32-bit
integers). These values are converted into text for display on the screen
using the format specifier for the front panel display. Normally, the
specifier used to parse a particular datum from the input stream will be
strongly related to the specifier used to display it (such as, all of the floating
point inputs will be displayed with an 'f' output specifier, and all of the
integer inputs will be displayed with a 'd' specifier).
The first line of the layout response is the scanf-like parameter list for
parsing the fields from an ASCII erec response. Parameters are separated by
spaces and the line is terminated by a “\n” (the normal line separator
character). Valid fields are:
%s - parse a string
%d - parse a decimal number
%ld - parse a long (32-bit) decimal number
%f - parse a floating point number
%x - parse a hexadecimal number
%lx - parse a long (32-bit) hex number
%* - ignore the field
Note
Signed versus unsigned for the integer values does not matter; it is
handled automatically.
▲
The second line of the layout response is the binary parameter list for
parsing the fields from a binary response. Parameters MUST be separated
by spaces, and the line is terminated by a '\n'. Valid fields are:
t - parse a time specifier (2 bytes)
D - parse a date specifier (3 bytes)
i - ignore one 8-bit character (1 byte)
e - parse a 24-bit floating point number (3 bytes: n/x)
Record Layout
Definition
Format Specifier for
ASCII Responses
Format Specifier for
Binary Responses
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......