Operation
Service Menu
3-100
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
ANALOG OUTPUT CAL: ZERO
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 100
SAVE VALUE
INC/DEC
SET OUTPUT TO: 0.0 V
RANGE AVG DIAGS ALARM
The Analog Output Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog output. The operator must connect
a volt meter to the output and adjust the output until the meter reads the
value shown in the “set output to:” field, in either V or mA, depending on
the selected output channel.
●
In the Main Menu, choose Service > Analog Out Cal > select
Channel >
Calibrate Full Scale
.
ANALOG OUTPUT CAL: SPAN
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 3397
SAVE VALUE
INC/DEC
SET OUTPUT TO: 10 V
RANGE AVG DIAGS ALARM
The Analog Input Calibration menu is used to calibrate the 8 analog input
channels at both zero and full-scale. The analog input calibration menu is
visible only when the I/O expansion board is installed and when the
instrument is in service mode. For more information on the service mode,
see “
Service Mode
” earlier in this chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Analog Input Calibration
.
Analog Output Calibrate
Full-Scale
Analog Input Calibration
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......