Calibration
Calibration Procedure
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
4-11
●
●
●
●
The span foil calibration procedure is now active and after another 270
seconds the calibration procedure should end, whereby a new Mass
Coef value is shown and the percent difference is shown from the last
foil calibration to the current calibration. This percent difference can
also be used to directly audit the mass coefficient value since a percent
difference in a mass coefficient is the same percent mass difference
overall.
MASS COEF: 6878.7
FOIL VALUE: 132
8
3
μ
g
Beta Avg 9911 1/SEC
difference -1.7 %
REMOVE SPAN FOIL
TO CONTINUE
RANGE AVG DIAGS ALARM
6.
After the span calibration, the screen will ask you to press
to
accept the new calibration. Otherwise press
to keep the old Mass
Coef value. Remove the span foil and press
. The bench will now
open. Remove the filter holder, replace the filter tape and press
to confirm the calibration routine has now been competed.
MASS COEF: 6878.7
FOIL VALUE: 132
8
3
μ
g
Beta Avg 9911 1/sec
difference -1.7 %
REMOVE HOLDE, INSERT
TAPE,
TO CLOSE BENCH
RANGE AVG DIAGS ALARM
MASS COEF: 7000.0
FOIL VALUE: 132
8
3
µ
g
Beta Avg 12022 1/sec
INSERT SPAN FOIL
TO START MASS CAL
RANGE AVG DIAGS ALARM
MASS COEF: 7000.0
FOIL VALUE: 132
8
3
µ
g
Beta Avg 9884 1/sec
PLEASE WAIT 245 sec
RANGE AVG DIAGS ALARM
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......