6.3 Measurement examples
90
*
only when selecting a network B1 or B2 (B1 selected, patient leakage current I)
Measuring the class-I equipment and class-II equipment
1.
Connect the instrument and the equipment under test.
See
"3.3 Connecting equipment under test to the instrument" (p. 36)
2.
Connect a test lead to terminal T2.
3.
Position the tip of the test lead on the applied part of the equipment under test.
4.
During manual measurement, patient leakage current (patient connection - earth)/patient leak-
age current I measurement starts upon return to the measurement screen.
Measurement conditions can be changed during measurement.
See
"6.1 Making manual measurements" (p. 79)
During automatic measurements,
becomes available upon return to the measurement
screen. Press
to start patient leakage current (patient connection - earth)/patient leakage
current I measurement.
Only
is available during automatic measurements.
Patient leakage current measurement
*
(Patient connection - Earth)
1
2
3
Terminal T2 and applied part
4
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......