8.7 Command Message Description
191
8
Chapter 8
Contro
lling the
Instru
ment From
a Comp
uter
Sets and queries the measurement network filter
Syntax
Command
Query
Response
:CONFigure:FILTer
<characters>
:CONFigure:FILTer?
<characters>
<characters>
= ON/ ON1_U2/ ON2_U3/ ON1_U1/ON2_U1/OFF
When network A is selected
OFF
: Single-frequency network
ON
: Multi-frequency network
When network B1 or B2 is selected(only ST5540)
ON
: Network with a frequency characteristic
OFF
: Network with 1 k
non-inductive resistance only
When network D is selected
ON1_U2
: Network compatible with perception/reaction U2
ON2_U3
: Network compatible with abandonment U3
ON1_U1
: Network compatible with perception/reaction U1
ON2_U1
: Network compatible with abandonment U1
OFF
: Body impedance network
When network D is selected
OFF
: 1.5 k
//0.15
F network
When network E is selected
OFF
: 1 k
network
When network F is selected
OFF
: 2 k
network
When network G is selected
OFF
: Network for damp conditions
Function
Command
Sets the measurement network filter.
Query
Returns the setting of the measurement network filter as alphabetic characters.
Example
Command
:CONFigure:FILTer OFF
Sets the network only with 1 k
non-inductive resistance.
(When network B2 is selected)
Query
Response
:CONFigure:FILTer?
:CONFIGURE:FILTER OFF
(Headers: ON)
OFF
(Headers: OFF)
The network only with 1 k
non-inductive resistance has been set.
(When network B2 is selected)
Error
An execution error occurs in the following cases:
• When the measurement mode is not selected
• When a mode other than "leakage current mode" is set
• In automatic measurement (Command)
Note
• Use this command after executing the
:MODE
command to activate the mea-
surement mode.
Sets and queries the measuring time of automatic measurement
Syntax
Command
Query
Response
:CONFigure:MTIMe
<numerical value>
:CONFigure:MTIMe?
<numerical value>
<numerical value>
= 1 to 300 (NR1)
Function
Command
Sets the measuring time of automatic measurement.
Query
Returns the measuring time of automatic measurement as a numeric value.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......