8.7 Command Message Description
219
8
Chapter 8
Contro
lling the
Instru
ment From
a Comp
uter
Error
I
f the set data unit number is higher than the total number of data units, an exe-
cution error occurs.
Reads saved data
Maxi-
mum
value
Judge
ment
Power
supply
polarity
Equipment under
test status
Filter
Target
current
Application
of 110% of
rated
voltage
under
other
conditions
Specific
110%
voltage
applica-
tion
Switch
2.345
mA
PASS
Positive
phase
Normal condition
ON
AC+DC
None
None
All Off
2.362
mA
PASS
Negative
phase
Normal condition
ON
AC+DC
None
None
All Off
2.510
mA
PASS
Positive
phase
Single-fault condition
(
Disconnection of protective earth
conductor
)
ON
AC+DC
None
None
All Off
2.610
mA
FAIL
Negative
phase
Single-fault condition
(
Disconnection of protective earth
conductor
)
ON
AC+DC
None
None
All Off
2.456
mA
PASS
Positive
phase
Single-fault condition
(Disconnection of one wire in
power line)
ON
AC+DC
None
None
All Off
Saves automatic measurement results
Syntax
Command
:MEMory:SAVE:AUTO
Function
Command
Saves automatic measurement results(date, Equipment name, Control number,
Grounding class, Applied part, network, measurement mode, Maximum value,
test condition setting at maximum value acquisition, judgement result)in mem-
ory.
*
When network B1 or B2 is selected
Example
Command
:MEMory:SAVE:AUTO
Saves automatic measurement results.
Error
An execution error occurs in the following cases:
• When the measurement mode is not selected
• When a mode other than "leakage current mode" is set
• In automatic measurement
• When there is not enough memory space
• When there is not enough memory space
• When the equipment name and control number are the same as those of
saved data, but the network, grounding class, and applied part differ
• When there is no data
• In manual measurement
Note
• Use this command after executing the
:MODE
command to activate the mea-
surement mode.
• After automatic measurement, save data before starting the next automatic
measurement.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......