9.5 Timing Chart
252
The following shows the output signal timing chart for automatic measurement.
The number of measurement items (n) varies depending on [number of polarity conditions] x [num-
ber of equipment status] set with
displayed on the automatic measurement setting screen.
(Example)
Polarity
: Positive polarity ON, negative polarity ON......2
Equipment status : Normal condition ON,
earth conductor disconnection ON ................2
With the above settings, the number of measurement items (n) is 4 (2 x 2 = 4).
*1 : Judgement result of item 1
*2 : Judgement result of item 2
*3 : Judgement result of item 3
*4 : Judgement result of item (n - 1)
*5 : Judgement result of item n
Description
Time
MIN
MAX
T1
Measurement start signal pulse width
1 ms
–
T2
Automatic measurement time
2 s
(T3 + T4) x number of measurement items
T3
Wait time between setting items
1 s
According to measurement delay setting for
automatic measurement
T4
Measuring time of each setting item
1 s
According to measuring time setting for auto-
matic measurement
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......