10.5 Other Functions
262
Maximum value hold
function
Effective in all measurement modes
Reset by Clear key during sampling
Reset by changing allowable value
Reset by START signal (EXT I/O, communication command)
Allowable value
judgement function
Allowable value : Sets the upper-limit current value/lower-limit current value.
Judgment :
PASS
Lower limit value measurement value Upper limit value
FAIL
Measurement value > upper-limit current value
LOW
Measurement value < lower-limit current value
"---"
Measurement error
(No measurement value can be obtained or instrument is
outside guaranteed accuracy range)
Processing : Indication, beep, judgment output from EXT I/O
Beep
: Selectable from "ON when PASS," "ON when FAIL," and "OFF"
Mode selection
function
(a) Current measurement
function
(1) Unit of current measurement : Auto
(2) Unit of current measurement : mA, fixed
Normal leakage current measurement function,
selectable unit of current measurement (default)
(b) Voltage
measurement
function
Isolates the internal network for using the instru-
ment as a voltmeter between terminals T1 and T2
Selectable target voltage and measurement range
Maximum measurement voltage : 50 V
(c) Protective conductor
current measurement
Selectable modes (AC, DC, AC + DC, ACPeak)
Setting the grounding
class and applied part
of equipment under
test
Selected from Class-I equipment, Class-II equipment, or internally powered
equipment
(When selecting network B1 or B2, select the applied part from the Type B
applied part, Type BF applied part, or Type CF applied part)
*
*
only ST5540
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......