8.7 Command Message Description
223
8
Chapter 8
Contro
lling the
Instru
ment From
a Comp
uter
Error
An execution error occurs in the following cases:
• When a mode other than "leakage current mode" is set (Command)
• When network is OFF(Command)
• During automatic measurements(Command)
• When power on polarity switching is set and a power on check ends in an
error (NG), it is the instrument that is causing the error.
• When network A, C, D, E, or F is selected
When Earth leakage current measurement mode is set for equipment under
test whose grounding class is not Class I
When set to Patient leakage current I, Patient leakage current II, Patient leak-
age current III, Patient auxiliary current, Patient leakage current Patient con-
nection - Earth, Patient leakage currentSIP/SOP, Patient leakage current F-
type applied part, Patient leakage current Metal accessible part, Total patient
leakage current Patient - Earth, Total patient leakage current SIP/SOP, Total
patient leakage current F-type applied part, Total patient leakage current
Metal accessible part mode
• When an attempt is made to set enclosure - line leakage current mode and
touch current (Enclosure - Line) when the grounding class of equipment
under test is internally powered equipment
• When network B1 is selected (only ST5540)
When Earth leakage current measurement mode is set for equipment under
test whose grounding class is not Class I
When patient leakage current III mode is set for equipment under test whose
applied part is set to Type B
When patient leakage current II mode is set for a Type BF or Type CF applied
part
When
Enclosure - Line leakage current is set
• When network B2 is selected(only ST5540)
When Earth leakage current measurement mode is set for equipment under
test whose grounding class is not Class I
When an attempt is made to set patient leakage current F-type applied part or
total patient leakage current F-type applied part for equipment under test
whose applied part is set to a B-type applied part
When an attempt is made to set patient leakage current metal accessible part
or total patient leakage current metal accessible part for a CF-type applied
part
Sets and queries the measurement mode
Sets and queries the network
Syntax
Command
Query
Response
:NETWork
<characters>
:NETWork?
<characters>
<characters>
= A/ B2/ C/ D/ E/ F/ G/B1/OFF
A: Network A
B2: Network B2(only ST5540)
C: Network C
D: Network D
E: Network E
F: Network F
G: Network G
B1: Network B1(only ST5540)
OFF: No network setting
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......