8.7 Command Message Description
187
8
Chapter 8
Contro
lling the
Instru
ment From
a Comp
uter
Sets and queries the allowable value (lower limit) On/Off
(when making AC measurements)(only ST5540)
Syntax
Command
Query
Response
:CONFigure:COMParator:LOWerAC
<character value 1>,<character value 2>
:CONFigure:COMParator:LOWerAC?
<character value 1>,<character value 2>
<character value 1>
= OFF/ON
OFF : Allowable value data for normal condition (lower limit : AC mea-
surements)OFF
ON : Allowable value data for normal condition (lower limit : AC measure-
ments)ON
<character value 2>
= OFF/ON
OFF : Allowable value for fault condition (lower limit : AC measure-
ments)OFF
ON : Allowable value for fault condition (lower limit : AC measure-
ments)ON
Function
Command
Sets the setting of On/Off of an allowable value (lower limit: AC measure-
ments).
Sets one On/Off setting of an allowable value (lower limit: during AC measure-
ments) each for normal and single fault conditions.
For internally powered equipment, set the On/Off setting of an allowable value
(lower limit : during AC measurements) for fault condition to Off.
Query
Returns the setting of On/Off of an allowable value (lower limit: AC measure-
ments).
Returns one On/Off setting of an allowable value (lower limit : during AC mea-
surements) for each normal and single fault conditions.
Internally powered equipment
Returns Off for the set of On/Off setting of the allowable value (lower limit: dur-
ing AC measurements) for fault condition.
Example
Command
:CONFigure:COMParator:LOWerAC ON,ON
Sets the On/Off setting of an allowable value (lower limit : during AC measure-
ments) for normal condition to On and the On/Off setting of an allowable value
(lower limit: during AC measurements) for single fault condition to On.
Query
Response
:CONFigure:COMParator:LOWerAC?
:CONFIGURE:COMPARATOR:LOWERAC ON,ON
(Headers: ON)
ON,ON
(Headers: OFF)
The On/Off setting of the allowable value (lower limit) for normal condition dur-
ing AC measurements is set to On and the On/Off setting of the allowable value
(lower limit) for single fault condition is set to On.
Error
An execution error occurs in the following cases:
• When patient auxiliary current, patient leakage current I, patient leakage cur-
rent (patient - earth), patient leakage current (SIP/SOP), total patient leakage
current (patient - earth) or total patient leakage current (SIP/SOP) measure-
ment mode is not selected.
• When a mode other than "leakage current mode" is set
• In automatic measurement (Command)
Note
Use this command after using the
:MODE
command to set one of the following
measurement modes: patient auxiliary current or patient leakage current I,
patient leakage current (patient - earth), patient leakage current (SIP/SOP),
total patient leakage current (patient - earth) or total patient leakage current
(SIP/SOP).
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......