5.1 Connecting the Test Lead
76
When selecting network B1
When selecting network B2
Class-I equipment
Class-II equipment
Internally powered equip-
ment
Type B Type BF Type CF Type B Type BF Type CF Type B Type BF Type CF
Earth leakage current
–
–
–
Enclosure - Earth
leakage current
Normal
T2
T2
T2
T2
T2
T2
T2
T2
T2
Fault/
Apply
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
Enclosure - Enclosure
leakage current
Normal
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
Fault/
Apply
T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3
Patient auxiliary current
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
Patient leakage current I
T2
T2
T2
T2
T2
T2
T1,T2
T1,T2
T1,T2
Patient leakage current II
T2,T3
–
–
T2,T3
–
–
T2,T3
–
–
Patient leakage current III
–
T2
T2
–
T2
T2
–
T2
T2
Free current
Normal
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
Fault/
Apply
T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3
Class-I equipment
Class-II equipment
Internally powered
equipment
Type B Type BF Type CF Type B Type BF Type CF Type B Type BF Type CF
Earth leakage current
–
–
–
Touch current
(Enclosure - Earth)
Normal
T2
T2
T2
T2
T2
T2
T2
T2
T2
Fault/
Apply
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
Touch current
(Enclosure - Enclosure)
Normal
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
Fault/
Apply
T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3
Patient auxiliary current
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
Patient leakage current
(Patient connection - Earth)
T2
T2
T2
T2
T2
T2
T1,T2
or T2
T1,T2
or T2
T1,T2
or T2
Patient leakage current
(external voltage on a SIP/SOP)
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
Patient leakage current
(external voltage on a specific
F-type applied part)
–
T2
T2
–
T2
T2
–
T2
T2
Patient leakage current
(external voltage on metal accessi-
ble part not protectively earthed)
T2,T3
T2,T3
–
T2,T3
T2,T3
–
T2,T3
T2,T3
–
Total patient leakage current
(Patient connection - Earth)
T2
T2
T2
T2
T2
T2
T1,T2
T1,T2
T1,T2
Total patient leakage current
(external voltage on a SIP/SOP)
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
Total patient leakage current
(external voltage on a specific
F-type applied part)
–
T2
T2
–
T2
T2
–
T2
T2
Total patient leakage current
(external voltage on metal accessi-
ble part not protectively earthed)
T2,T3
T2,T3
–
T2,T3
T2,T3
–
T2,T3
T2,T3
–
Free current
Normal
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
Fault/
Apply
T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3 T1,T2,T3
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......