10.4 Accuracy
259
8
9
Chap
ter 10
S
pe
c
ificat
ions
10
Protective conductor current accuracy
Measurement of DC
Measurement of AC/AC + DC
(During AC measurements, the frequency characteristics of the high-pass filter (fc=4 Hz) is added)
ACPeak
Network accuracy
Range
Guaranteed accuracy range
Measurement resolution
Accuracy
50.00 mA
12.00 mA to 50.00 mA
10
A
± 2.0%rdg.±6dgt.
10.00 mA
1.30 mA to 13.00 mA
10
A
± 2.0%rdg.±6dgt.
Range
Guaranteed
accuracy range
Measurement
resolution
Accuracy
15 Hz f 100 kHz
100 kHz < f 1 MHz
50.00 mA
12.00 mA to 50.00 mA
10
A
± 2.0%rdg.±6dgt.
± 5.0%rdg.±20dgt.
10.00 mA
1.30 mA to 13.00 mA
10
A
± 2.0%rdg.±6dgt.
± 5.0%rdg.±20dgt.
Range
Guaranteed
accuracy range
Measurement
resolution
Accuracy
15Hz
f 10 kHz
10 kHz <
f
100 kHz
100 kHz < f
1 MHz
75.0 mA
12.0 mA to 75.0 mA
100
A
± 2%rdg.±6dgt.
± 5.0%f.s.
± 25.0%f.s.
10.00 mA
1.30 mA to 13.00 mA
10
A
± 2.5%f.s.
± 5.0%f.s.
± 25.0%f.s.
Network Section
Network name/filter status
Characteristic*1, *2
DC input resistance
Frequency range
with
±
1% deviation
Cut-off frequency
(-3 dB points*4)
A/Filter OFF
*3
1 k
±1%
100 kHz or lower
–
A/Filter ON
1 k
±1%
–
1326 ±20 Hz
B1
*
, B2
*
/ Filter OFF
*3
1 k
±1%
100 kHz or lower
–
B1
*
, B2
*
/ Filter ON
1 k
±1%
–
1047
±16 Hz
C/Filter OFF
2 k
±1%
–
1811
±27 Hz
C/Filter ON1
2 k
±1%
–
3470
±104 Hz
*4
C/Filter ON2
2 k
±1%
–
9100
±273 Hz
*4
D
1.5 k
±1%
–
705
±15 Hz
E
*3
1 k
±1%
100 kHz or lower
–
F
2 k
±1%
100 kHz or lower
–
G
875
±1%
–
1997 Hz±27 Hz
*
only ST5540
Network
name/filter
status
Accuracy (Deviation from theoretical value. Includes accuracy of internal voltmeter)
Measured current
AC, AC+DC
ACpeak
range
50 mA, 5 mA,
500
A
50
A
75 mA, 10 mA
1 mA, 500
A
A/Filter
OFF
*3
0.1Hz <
f <
10 kHz
10 kHz f
< 100 kHz
100 kHz
f
1MHz
± 4%rdg.±10dgt.
± 4%f.s.
–
–
A/Filter
ON
0.1Hz <
f <
10 kHz
10 kHz f
< 100 kHz
100 kHz
f
1MHz
± 4%rdg.±10dgt.
± 1.5dBrdg.±10dgt.
± 1.5dBrdg.±10dgt.
± 4%f.s.
± 1.5dBrdg.±2%f.s.
± 1.5dBrdg.±2%f.s.
–
–
B1
*
, B2
*
/
Filter
OFF
*3
0.1Hz <
f <
10 kHz
10 kHz f
< 100 kHz
100 kHz
f
1MHz
± 4%rdg.±10dgt.
± 4%f.s.
–
–
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......