8.7 Command Message Description
222
When network B2 is selected
OFF
: Mode not selected
(Initial screen : command)
(Initial screen, system screen: query)
EARTh
: Earth leakage current
TOUCh1
: Touch current Enclosure - Earth
TOUCh2
: Touch current Enclosure - Enclosure
PAUXiliary
: Patient auxiliary current
PATientP2E
: Patient auxiliary current Patient - Earth
PATientSIPSOP
: Patient leakage current SIP/SOP
PATientFTYPE
: Patient leakage current F-type applied part
PATientMP
: Patient leakage current Metal accessible part
TPATientP2E
: Total patient leakage current
Applied part- Earth
TPATientSIPSOP : Total patient leakage current SIP/SOP
TPATientFTYPE
: Total patient leakage current F-type applied part
TPATientMP
: Total patient leakage current Metal accessible part
FREE
: Free current
Function
Command
Sets the measurement mode and changes the screen.
Query
Returns the measurement mode setting as alphabetic characters.
Example
Command
:MODE EARTH
Sets Earth leakage current measurement mode.
Query
Response
:MODE?
:MODE EARTH
(Headers : ON)
EARTH
(Headers : OFF)
Earth leakage current measurement mode has been set.
Sets and queries the measurement mode
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......