4.6 Setting the Allowable Value
61
3
Cha
pter 4
Se
ttings
4
Set the allowable value (upper-limit leakage current value) for the measurement value and maximum
value.
According to the status (normal condition, single-fault condition) of the equipment under test, two values
can be set.
The allowable value setting is based on "Numeric value x Factor."
Use the allowable value setting to provide a margin or compensate for measurement error.
The factor is set to "100%" by default. (Input range 1 to 100% )
Set a lower limit value when required.
• Setting a lower limit value for all measurement modes.
See
"7.9 Setting Lower Limit Values (factor)" (p. 130)
• Setting a lower limit value for specific measurement modes.
See
"Turning lower limit values on and off for specific leakage current measurements" (p. 63)
The instrument indicates the result (PASS/FAIL/LOW indication) of comparison between the measure-
ment value and allowable value.
In automatic measurement, the judgement result is output from EXT I/O.
See
"9.2 Connecting to the EXT I/O Terminal" (p. 247)
The buzzer that sounds based on the judgement result can be disabled.
See
"7.13 Setting the Beep Sound" (p. 135)
4.6 Setting the Allowable Value
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......