5.1 Connecting the Test Lead
77
5
Chapt
er 5 Measu
reme
nt Prep
arations
Connect the L2200 Test Lead to a measuring terminal (T1, T2, or T3).
The terminal to be used varies depending on the measurement mode.
See
"5.1 Connecting the Test Lead" (p. 75)
When measuring enclosure leakage current, position the surface contact probe on the enclosure of the
equipment under test, then set the test lead on the surface contact probe.
Ensure full contact of the surface contact probe on the enclosure by applying about 0.5 N/cm2 of pres-
sure.
Using L2200 Test Lead
To avoid the risk of electric shock, do not touch the tips of test leads con-
nected to terminals T1, T2, and T3.
In some measurement modes, high voltage is output from the terminals.
Do not touch any part of
the test lead beyond the
finger barrier
To apply 110% of rated voltage, use an isolation transformer to supply the volt-
age to [LINE IN].
The voltage applied to [LINE IN] is also output from the T3 terminal.
Using the 9195 Enclosure Probe
(for measuring enclosure leakage current)
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......