1.6 Types of Leakage Current Measurement
15
1
Chapter 1
Overview
Patient auxiliary current is current that flows through an "applied part human body applied part" path.
Regardless of the medical equipment class, type of applied parts, and signal input/output selection, all
medical equipment with multiple applied parts must undergo this measurement.
Make both AC and DC measurements for this type of leakage current.
This is electric current that flows through an "
applied part
human body
ground" path.
Regardless of medical equipment class, type of
applied parts
and
signal input/output section
, all medical
equipment with
patient connections
must undergo this measurement.
Make both AC and DC measurements for this type of leakage current.
Patient auxiliary current
Patient leakage current measurement (Patient connection - Earth)
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......