8.7 Command Message Description
214
Error
If the measurement mode is not set, an execution error occurs.
When there is no maximum value, +9.999E+10,3,0,0,0,0,0 will be displayed.
Note
• To obtain the maximum value acquired in automatic measurement, execute
this command after confirming the completion of automatic measurement
(
:AMC?
command). If the command is executed during automatic measure-
ment, the maximum value at command execution is obtained.
• Use this command after executing the
:MODE
command to activate the mea-
surement mode.
Queries the maximum value
Maximum
value
Judge
ment
Power
supply
polarity
Equipment under test status
Target
current
Other 110%
voltage
application
Specific
110% volt-
age applica-
tion
2.345 mA
FAIL
Negative
phase
Single-fault condition(Disconnec-
tion of protective earth conductor)
AC+DC
none
none
Queries the protective conductor current measurement value
Syntax
Query
Response
:MEASure:PCC?
<numerical value>
<numerical value>=
Measurement value
(NR3)
Function
Example
Query
Returns protective conductor current measurement value as a numeric value.
(unit : A)
Query
:MEASure:PCC?
Error
Response
:MEASURE:PCC +2.345E-03
(Headers : ON)
+2.345E-03
(Headers : OFF)
Protective conductor current measurement value is 2.345 mA.
Mode is set to something other than "protective conductor ammeter mode"
Note
• Use this command after using the
:SYSTem:MODE
command to set the
protective conductor ammeter mode.
Queries the voltage measurement value
Syntax
Query
Response
:MEASure:VOLTage?
<numerical value>
<numerical value>
= Measurement value (NR3)
Function
Query
Returns the voltage measurement value as numeric values.(unit : V)
Example
Query
Response
:MEASure:VOLTage?
:MEASURE:V2.345E+00
(Headers : ON)
+2.345E+00
(Headers : OFF)
The voltage measurement value is 2.345 V
Error
An execution error occurs when mode is set to something other than "voltmeter
mode."
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......