7.7 Power On Polarity Switching (to keep equipment under test on during polarity
128
Relay check 2
Do not connect the power cord to the terminal block on the rear panel.
Since turning the relay On supplies power from front terminal block and
the outlet, a power cord connection to the rear terminal block can cause
electric shock.
•
Perform a check without connecting equipment under test.
•
When the impedance value is
or OVER FLOW and the relay is Off, the fuse
in the instrument might have blown.
Performing a fuse check may lead to recovery. (p. 123)
1.
Set the instrument circuit breaker to On.
2.
Set the ON/OFF relay to
.
3.
Use a tester or other measuring instrument to
measure the impedance value between the T1
terminal and the L terminal on the terminal block
on the rear panel of this instrument.
4.
Make sure that the impedance value is approxi-
mately 102 k
.
5.
Set the ON/OFF relay to
.
6.
Like step
2
, use a tester or other measuring
instrument to measure the impedance value
between the T1 terminal and the L terminal on the
terminal block on the rear panel of this instru-
ment.
7.
Make sure that the impedance value is
or
OVER FLOW.
Required items : A device measuring impedance
(for example, a tester)
2
5
Relay check 2 ON/OFF setting screen.
E N L
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......