11.4 Instrument Disposal
283
11
Chapte
r 1
1
Ma
inten
ance
and Service
The instrument contains a lithium battery for system memory backup.
11.4 Instrument Disposal
• To avoid electric shock, turn off the power switch and disconnect the
power cord and test leads before removing the lithium battery.
•
When disposing of this instrument, remove the lithium battery and dis-
pose of battery and instrument in accordance with local regulations.
• Keep batteries away from children to prevent accidental swallowing.
If the protective functions of the instrument are damaged, either remove it from
service or mark it clearly so that others do not use it inadvertently.
Removing the lithium battery
1.
Turn the power switch OFF (
), then discon-
nect the power cord and test leads.
2.
Remove the 8 screws that secure the legs on
the rear panel, and the 2 screws used to
mount the handle on the side panel.
3.
Remove the top case section by sliding it to
the back of the instrument.
4.
Locate the battery installed at the corner of the
circuit board. Insert the tip of forceps between
the battery holder and battery to lift the battery
for removal.
Required tools
• Phillips screwdriver
1 pc.
• Forceps
1 set
CALIFORNIA, USA ONLY
Perchlorate Material - special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......