8.7 Command Message Description
207
8
Chapter 8
Contro
lling the
Instru
ment From
a Comp
uter
Sets and queries the equipment name/control number of equipment under test
Syntax
Command
Query
Response
:EQUipment:IDENtity
<character value 1>,<character value 2>
:EQUipment:IDENtity?
<character value 1>,<character value 2>
<character value 1>
= Equipment name
(text data with 1 to 12 characters)
<character value 2>
= Control number
(text data with 1 to 12 characters)
Function
Command
Sets the equipment name/control number of equipment under test.
Query
Returns the equipment name/control number setting of equipment under test.
Example
Command
:EQUipment:IDENtity ABC,NO-111
Sets the equipment name/control number of equipment under test to "ABC" and
"NO-111," respectively.
Query
Response
:EQUipment:IDENtity?
:EQUIPMENT:IDENTITY ABC,NO-111
(Headers : ON)
ABC,NO-111
(Headers : OFF)
The equipment name/control number of equipment under test have been set to
"ABC" and "NO-111," respectively.
Error
An execution error occurs in the following cases:
• When the measurement mode is selected(Command)
• When a mode other than "leakage current mode" is set(Command)
• If the character string does not conform to the specified format (Command)
Note
• Use this command after executing the
:MODE
command to activate OFF
mode (i.e., measurement mode not selected).
• Alphanumeric characters and hyphens (-) can be used. All letters are treated
as uppercase (not case-sensitive).
The equipment name and control number need not be set except when mea-
surement data is saved in memory.
Sets and queries the applied part of equipment under test
(Network B1, B2 only)(only ST5540)
Syntax
Command
Query
Response
:EQUipment:TYPE
<characters>
:EQUipment:TYPE?
<characters>
<characters>
= B/ BF/ CF
B : Type B applied part
BF : Type BF applied part
CF : Type CF applied part
Function
Command
Sets the applied part of equipment under test.
Query
Returns the applied part setting of equipment under test.
Example
Command
:EQUipment:TYPE B
Sets the applied part of equipment under test to Type B applied part.
Query
Response
:EQUipment:TYPE?
:EQUIPMENT:TYPE B
(Headers: ON)
B
(Headers: OFF)
The applied part of equipment under test has been set to Type B applied part.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......