1.6 Types of Leakage Current Measurement
19
1
Chapter 1
Overview
Total patient leakage current is the total of all leakage current to or from
patient connections
of all the
same type of
applied parts
(
B-Type, BF-Type and CF-Type applied parts
) simultaneously connected.
When there are two or more
patient connections
that go to different functions and are not electrically con-
nected together, the total patient leakage current for
B-Type applied parts
must be measured.
If required, measure all patient leakage current components (
patient connection
- earth, external voltage
on an SIP/SOP, external voltage on an
F-type applied part
, external voltage on a metal accessible part
that is not protectively earthed.)
(The figure shows patient - earth.)
Total patient leakage current measurement