8.7 Command Message Description
164
Restores the default settings (factor settings)
Syntax
Command
*RST
Function
Command
I
nitializes the instrument settings.
The initial screen appears after initialization is completed.
Same as specifying
ALL
with HiTESTER initialization command
:SYS-
Tem:RESet
Example
Command
*RST
Initializes the instrument settings.
Error
Any data following a command results in a command error.
Note
The communication conditions are not initialized.
Queries the self-test and result
Syntax
Query
Response
*TST?
<numerical value>
<numerical value>
= 0, 2 (NR1)
0: No error
2: RAM error
Function
Query
Conducts a self-test of main instrument and returns the result as a numeric
value of 0 or 2.
Example
Query
Response
*TST?
2
A RAM error has occurred.
Error
If the response message exceeds the output queue size, a query error occurs.
Note
A self-test initiated by a communications command is a RAM test only. The key
test, LCD test, (zzz) test, and LED test on the self-test screen are not con-
ducted by a communications command.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......