8.5 Initialization Items
153
8
Chapter 8
Contro
lling the
Instru
ment From
a Comp
uter
Event status registers specific to the HiTESTER (ESR0)
An event status register has been provided to manage instrument events.
The event status registers are each 8-bit registers.
It is possible to determine the status of the instrument by reading these regis-
ters.
The data in event status register 0 is cleared under the following conditions.
•
When a
*CLS
" command is received.
•
When an
:ESR0?
query (for event status register 0) is received.
When the instrument is powered on.
*1 Except the PON bit (bit 7).
Event Status Register 0 (ESR0)
Bit 7
Not used
Bit 6
ERROR
Judgment Judgment fail
Bit 5
LOW
When the maximum value goes below the lower
limit value
Bit 4
TEST
In automatic measurement
Bit 3
MEAS
Performing measurement under each condition
Bit 2
T-FAIL
Total failure(Automatic measurement only)
Bit 1
FAIL
Failure
Bit 0
PASS
Pass
8.5 Initialization Items
Initialization method
Item
Power on
*RST
command
*CLS
command
Device specific functions (ranges etc.)
Output queue
Input buffer
Event registers
*
1
Current path
Headers on/off
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......