8.7 Command Message Description
238
Initializes the ST5540/ST5541
Syntax
Command
:SYSTem:RESet
<characters>
<characters>
= ALL/ CONDition/ SAVEdata
ALL
:Initializes the instrument to the default
settings (factory settings).
(
Same as
*RST
command
)
CONDition :Initializes the instrument to the default settings except for
saved measurement data.
SAVEdata
:Initializes only the saved measurement data.
Function
Command
Initializes the HiTESTER.
After initialization, the initial screen is displayed.
Example
Command
:SYSTem:RESet SAVEdata
Initializes only the saved measurement data.
Note
The communication conditions are not initialized.
Saves a panel
Syntax
Command
:SYSTem:SAVE
<numerical value>
<numerical value>
= 1 to 30 (NR1)
Function
Command
Saves (stores) the present settings in the panel of the specified number.
Example
Command
:SYSTem:SAVE 5
Saves setting data in panel number 5.
Error
An execution error occurs in the following cases :
• When the measurement mode is not selected
• If the designated panel number does not conform to the specified range
• When a mode other than "leakage current mode" is set
Queries the fuse check
Syntax
Query
Response
:SYSTem:TEST:FUSE?
<characters>
<characters>
=OK/NG
OK: FUSE CHECK OK
NG: FUSE CHECK NG
Function
Query
Performs a fuse check and returns the result as alphabetic characters.
Example
Error
Query
Response
:SYSTem:TEST:FUSE?
:SYSTEM:TEST:FUSE OK
(Headers : ON)
OK
(Headers : OFF)
The result of the fuse check is OK.
An execution error occurs in the following cases :
• When measurement mode is selected
• When a mode other than "leakage current mode" is set
Note
• Use this command after using the
:MODE
command to turn off
(and no measurement mode is selected).
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......