8.7 Command Message Description
162
8.7 Command Message Description
Describes important notes
regarding use of the command.
"Command" is displayed only
when the note is related to command.
Queries power supply polarity setting
in manual measurements
Syntax Command
Query
Response
:CONFigure:POLarity
<characters>
:CONFigure:POLarity?
<characters>
<characters>
=NORMal/REVerse
NORMal:Positive phase
REVerse:Negative phase
Function Command
Sets the power supply polarity for manual measure-
ment.
Query
Returns the power supply polarity setting for manual
measurement as alphabetic characters.
Example Command
:CONFigure:AUTO OFF;:CONFigure:POLarity
NORMal
Sets the power supply polarity to positive phase.
Query
Response
:CONFigure:AUTO OFF;:CONFigure:POLarity?
:CONFIGURE:POLARITY NORMAL
(Headers: ON)
NORMAL
(Headers: OFF)
The power supply polarity has been set to positive
phase.
Error
An execution error occurs in the following cases:
• When the measurement method is set to "automatic
measurement"
• When a mode other than “leakage current mode” is
set
• When the measurement mode is set to Enclosure -
Line leakage current, Touch current Enclosure - Line
or the measurement mode is not set
• When the type of equipment under test is set to
"internally powered equipment"
Note
Use this command after executing the
:MODE
com-
mand to activate the measurement mode.
Shows the command message that contains numerical or character parameters.
<numerical value>
: Numeric data values
(NR1) integer data
(NR2) fixed point numbers
(NR3) floating point numbers
(NRf) format that contains NR1, NR2, and NR3
<characters>
: Character string data
<Data to be entered>
Shows the command description.
Describes the message syntax.
Explains the message.
Illustrates the actual command
application. (Generally, the explanation
is directed to the "HEADER ON" case,
except for the HEADER command.)
Describes errors that may occur
due to command execution.
"Command" is displayed only
when the error is related to command.
Command, Query
Response
ST5540 or ST5541
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......