8.7 Command Message Description
168
Numerical value1
Numerical value2
Example
Command
:CONFigure:AUTO:KIND 97,0
Sets the type of automatic measurement as follows:
Equipment under test status: Normal condition
Power supply polarity: Positive phase, Negative phase
Query
Response
:CONFigure:AUTO:KIND?
:CONFIGURE:AUTO:KIND 99,0
(Headers: ON)
99
(Headers: OFF)
The type of automatic measurement has been set as follows:
Equipment under test status: Normal condition, Disconnection of one wire in
power line
Power supply polarity : Positive phase
,
Negative phase
Error
An execution error occurs in the following cases:
• When the measurement method is set to "manual measurement"
• When the measurement mode is not selected
• If the set numeric value is not within the specified range(Command)
• When a mode other than "leakage current mode" is set
• In automatic measurement(Command)
Queries the type and setting of automatic measurement
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
Not used
Negative
phase
Positive
phase
Single-fault
condition
Application
of 110%
voltage
(
Negative
phase
)
Single-fault
condition
Application
of 110%
voltage
(
Positive
phase
)
Single-fault
condition
Disconnec-
tion of pro-
tective
earth con-
ductor
Single-fault
condition
Discon-
nection of
one wire in
power line
Normal
32768
16384
8192
4096
2048
1024
512
256
bit 15
bit 14
bit 13
bit 12
bit 11
bit 10
bit 9
bit 8
Not used
Not used
Not used
Not used
Application
of line
voltage N
Application
of line
voltage L
AC
DC
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
Not used
Not used
Specific
110%
voltage
application
(
Negative
phase
)
Specific
110%
voltage
application
(
Positive
phase
)
Not used
Other 110%
voltage
application
(
Negative
phase
)
Other 110%
voltage
application
(
Positive
phase
)
Other 110%
voltage
application
(none)
32768
16384
8192
4096
2048
1024
512
256
bit 15
bit 14
bit 13
bit 12
bit 11
bit 10
bit 9
bit 8
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......