6.4 Saving Measurement Data (As required)
105
6
Chapt
er 6 Measureme
nt
s
Selecting Saved Measurement Data
1.
Press
to switch the mode used for the data
unit.
On the sample screen, the mode changes in the fol-
lowing order:
Earth Leak
TC(En-En)
PL(Pa-Ea)
TPL(Pa-Ea).
2.
Press
to return to the system screen.
Measurement data reference screen
List of measurement modes (varies depending on network types)
Earth Leak
Earth leakage current
PL (Pa-Ea)
Patient leakage current
(Patient connection - Earth)
En-Ea Leak
Enclosure - Earth leakage current
PL-SIPSOP
Patient leakage current (SIP/SOP)
En-En Leak
Enclosure - Enclosure
leakage current
PL (F)
Patient leakage current (F-type applied part)
En-Li Leak
Enclosure - Line leakage current
PL (MP)
Patient leakage current
(Metal accessible part)
Pat.Leak I
Patient leakage current I
TPL (Pa-Ea)
Total patient leakage current
(Patient connection - Earth)
Pat.Leak II
Patient leakage current II
TPL-SIPSOP
Total patient leakage current (SIP/SOP)
Pat.Leak III
Patient leakage current III
TPL (F)
Total patient leakage current
(F-type applied part)
TC (En-Ea)
Touch current (Enclosure - Earth)
TPL (MP)
Total patient leakage current
(Metal accessible part)
TC (En-En)
Touch current (Enclosure - Enclosure)
Pat.Aux.
Patient auxiliary current
TC (En-Li)
Touch current (Enclosure - Line)
Free
Free current measurement
1
About data unit sorting
The sort function rearranges data unit numbers in sequence based on the equip-
ment name (1st character (A, B, C, ..., Z, 0, 1, 2, ..., 9, -), 2nd character, ..., 12th
character), and control number. When a new data unit is added, the data units
are sorted again.
Deleting Saved Data
1.
Use
and
to select a data unit to
delete.
2.
Press .
A confirmation screen appears.
3.
Press
to delete the selected data unit.
To delete two or more data units, repeat steps
1
through
3
.
4.
Press
to return to the system screen.
Measurement data reference screen
2
The initialization screen can be used to delete all saved data.
See
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......