10.14 Measurement Networks
272
Network C is a measurement network that complies with IEC 60990.
Network C complies with the following standards :
(1)
Methods of measurement of touch current and protective conductor current
(IEC 60990:1999)
Filter setting for human body impedance network (OFF), network compatible
with perception/reaction (ON1 (U1,U2)), and network compatible with abandon-
ment (ON2 (U1,U3))
When network C is selected, the instrument can also be used in compliance with
the following standards on testing leakage current :
(2)
Safety requirements for electrical equipment for measurement, control, and
laboratory use(IEC 61010-1:2010)
(3)
Safety of information technology equipment
(IEC 60950-1:2005)
Network C (for IEC 60990)
V
1.5 k
±
1%
0.22
F
±
2%
500
±
1%
0.0062
F
±
2%
20 k
±
1%
0.0091
F
±
2%
V
V
0.022
F
±
2%
0.22
F
±
2%
1.5 k
±
1%
500
±
1%
1.5 k
±
1%
0.22
F
±
2%
500
±
1%
10 k
±
1%
Network compatible with
perception/reaction(ON1(U2))
Human body impedance network(OFF)
Network compatible with
abandonment(ON2(U3))
10 k
±
1%
0.022
F
±
2%
1.5 k
±
1%
0.22
F
±
2%
500
±
1%
10 k
±
1%
V
Network compatible with perception/
reaction(ON1(U1))
V
1.5 k
±
1%
0.22
F
±
2%
500
±
1%
0.0062
F
±
2%
20 k
±
1%
0.0091
F
±
2%
Network compatible with
abandonment(ON2(U1))
10 k
±
1%
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......