9.5 Timing Chart
253
8
9
Chapte
r 9 Ext
ernal
Co
ntrol
Saved measurement conditions can be read and measurement conducted
according to the conditions read.
The number of the panel to be read is specified by LOAD0 through LORD4
before START signal input.
See
"Table of LOAD0 - LOAD4 control and corresponding panel numbers" (p. 248)
TEST
,
PASS
,
U-FAIL
,
L-FAIL
and
T-FAIL
are only output in automatic measure-
ment.
T5
Judgement result output time
2 s
Until measurement completion of next mea-
surement item
T6
Time from the completion of automatic
measurement to next measurement
start signal input
0 ms
–
Description
Time
MIN
MAX
Panel load function at EXT I/O start
Description
Time
Explanation
MIN
T7
From setting LOAD0 to LOAD4 to START
input
1 ms Low level must be held for at least 1 ms
before starting.
T8
From START input to changing the condition
in LOAD0 to LOAD4 (change of condition for
next measurement)
1 ms After START signal input, the panel load con-
dition must be held for at least 1 ms.
T9
From completion of automatic measurement
to next START input
0 ms When the condition is the same as T6, but
the panel to be loaded is changed for mea-
surement, it is necessary to change LOAD0
to LOAD4 in advance.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......