10.14 Measurement Networks
271
8
9
Chap
ter 10
S
pe
c
ificat
ions
10
This network complies with IEC 60601-1:2005 3rd edition.
The B2 measurement network complies with the following standard.
(1)
Medical electrical equipment - Part 1 : General requirements for safety
(IEC 60601-1:2005 3rd)
(2)
General safety regulations for medical electrical equipment (JIS T 1001-92)
(3)
Medical electrical equipment-Part1:
General requirements for basic safety and essential performace
(JIS T 0601-1:2012)
When network B2 is selected, the instrument can also be used in compliance
with the following standards on testing leakage current:
(4)
JIS standards for medical X-ray equipment
The instrument can be used for testing leakage current in isolation transformers
in compliance with the safety standard for electrical facilities for hospitals (JIS T
1022-96).
In addition to medical electrical equipment standards, many JIS standards spec-
ify the use of a network with 1 k
non-inductive resistance for testing leakage
current.
Network B2 (for IEC 60601-1:2005 3rd)
V
1 k
±
1%
10 k
±
1%
0.015
F
±
2%
V
1 k
±
1%
Network with a frequency
characteristic (ON)
Network with 1 k
non-inductive
resistance only (OFF)
Applicable
standards
(1) Medical electrical equipment - Part 1 : General requirements for safety
IEC 60601-1:2005 3rd, JIS T 0601-1:2012
(2) Medical electrical safety regulations
JIS T1001-92
(3) General regulations for method of testing safety of medical electrical equipment
JIS T1002-92
Measurement circuit
configuration
Basic measurement element: 1 k
Frequency characteristic: ±0.5% (DC to 1 MHz)
Filter configuration
Filter configuration (ON setting) : RC filter (10 k
+ 15 nF)
Filter setting
: ON/OFF (1 k
only at OFF)
(with short-circuited input protective fuse)
Element tolerance
Resistance : ± 1%
Capacitor : ± 2%
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......