10.4 Accuracy
260
*
1:
Measurement (including cable capacity) between T1 and T2 in leakage current between
enclosure and enclosure mode
*
2:
Including voltmeter (1 M
load) at network output section Input protective fuse is short-cir-
cuited.
*
3:
Network A (filter OFF), Network B1, B2 (filter OFF),Same circuit for network E
*
4:
-15 dB points for network C (filter ON1, ON2 )
*
5:
0.1 Hz only at Network B2. Other Network from 15 Hz.
*
6:
Impedance theoretical values do not include the network output unit voltmeter (1 M
load)
B1
*
, B2
*
/
Filter ON
0.1Hz <
f <
10 kHz
10 kHz f
< 100 kHz
100 kHz
f
1MHz
± 4%rdg.±10dgt.
± 1.5dBrdg.±10dgt.
± 1.5dBrdg.±10dgt.
± 4%f.s.
± 1.5dBrdg.±2%f.s.
± 1.5dBrdg.±2%f.s.
–
–
C/Filter
OFF
0.1Hz <
f <
10 kHz
10 kHz f
< 100 kHz
100 kHz
f
1MHz
± 4%rdg.±10dgt.
± 1.5dBrdg.±10dgt.
± 1.5dBrdg.±10dgt.
± 4%f.s.
± 1.5dBrdg.±2%f.s.
± 1.5dBrdg.±2%f.s.
–
–
C/Filter
ON1
0.1Hz <
f <
10 kHz
10 kHz f
< 100 kHz
100 kHz
f
1MHz
± 4%rdg.±10dgt.
± 1.5dBrdg.±10dgt.
± 3.5dBrdg.±10dgt.
± 4%f.s.
± 1.5dBrdg.±2%f.s.
± 3.5dBrdg.±2%f.s.
± 4%rdg.±10dgt.
± 1.5dBrdg.±5%f.s.
± 3.5dBrdg.±15%f.s.
± 4%f.s.
± 1.5dBrdg.±5%f.s.
± 3.5dBrdg.±15%f.s.
C/Filter
ON2
0.1Hz <
f <
10 kHz
10 kHz f
< 100 kHz
100 kHz
f
1MHz
± 4%rdg.±10dgt.
± 1.5dBrdg.±10dgt.
± 3.5dBrdg.±10dgt.
± 4%f.s.
± 1.5dBrdg.±2%f.s.
± 3.5dBrdg.±2%f.s.
± 4%rdg.±10dgt.
± 1.5dBrdg.±5%f.s.
± 3.5dBrdg.±15%f.s.
± 4%f.s.
± 1.5dBrdg.±5%f.s.
± 3.5dBrdg.±15%f.s.
D
*6
0.1Hz <
f <
10 kHz
10 kHz
f
1MHz
Theoretical
impedance value
± 2%
Including voltmeter
± 4%rdg.±6dgt.
Theoretical
impedance value
± 3%± 6
Including voltmeter
± 5%rdg.±6dgt.
Theoretical
impedance value
± 2%
Including voltmeter
± 4%rdg.±6dgt.
Theoretical
impedance value
± 3%± 6
Including voltmeter
± 5%rdg.±6dgt.
–
–
E
*3
0.1Hz <
f <
10 kHz
10 kHz f
100 kHz
100 kHz
f
1MHz
± 4%rdg.±10dgt.
± 4%f.s.
± 4%rdg.±10dgt.
±5%f.s.
±15%f.s.
± 4%f.s.
±5%f.s.
±15%f.s.
F
0.1Hz <
f <
10 kHz
10 kHz f
< 100 kHz
100 kHz
f
1MHz
± 4%rdg.±10dgt.
± 4%f.s.
± 4%rdg.±10dgt.
±5%f.s.
±15%f.s.
± 4%f.s.
±5%f.s.
±15%f.s.
G
0.1Hz <
f <
10 kHz
10 kHz f
< 100 kHz
100 kHz
f
1MHz
± 4%rdg.±10dgt.
± 1.5dBrdg.±10dgt.
± 1.5dBrdg.±10dgt.
± 4%f.s.
± 1.5dBrdg.±2%f.s
± 1.5dBrdg.±2%f.s
± 4%rdg.±10dgt.
± 1.5dBrdg.±5%f.s.
± 1.5dBrdg.±15%f.s.
± 4%f.s.
± 1.5dBrdg.±5%f.s.
± 1.5dBrdg.±15%f.s.
The value
x dBrdg is applied to the stipulated value (dB) under various stan-
dards.
Network
name/filter
status
Accuracy (Deviation from theoretical value. Includes accuracy of internal voltmeter)
Measured current
AC, AC+DC
ACpeak
range
50 mA, 5 mA,
500
A
50
A
75 mA, 10 mA
1 mA, 500
A
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......