8.7 Command Message Description
234
Sets and queries the panel name
Syntax
Command
Query
Response
:SYSTem:FILE:NAME
<numerical value>,<characters>
:SYSTem:FILE:NAME?
<numerical value>
<numerical value>
=1 to 30(NR1)
<characters>
<characters>
=panel name (Character data: 1 to 12 characters)
Function
Example
Command
Query
Sets the panel name of the specified number.
Returns the panel name setting of the specified number.
Command
:SYSTem:FILE:NAME 1,ABC
Sets the panel name of number 1 to "ABC."
Error
Query
Response
:SYSTem:FILE:NAME? 1
:SYSTEM:FILE:NAME ABC
(Headers: ON)
ABC
(Headers: OFF)
The panel name is set to "ABC."
If the designated panel number does not conform to the specified range, an
execution error occurs.
An execution error occurs when there is no data. (Command)
Note
Returns "0" when there is no data. (Query)
Sets and queries the frequency range
Syntax
Command
Query
Response
:SYSTem:FREQuency
<characters>
:SYSTem:FREQuency?
<characters>
<characters>
=15HZ/0.1HZ
15HZ:from
15Hz
0.1HZ:from 0.1Hz
Function
Example
Command
Query
Sets the frequency range.
Returns the frequency range setting as alphabetic characters.
Command
:SYSTem:FREQuency 15HZ
Sets the frequency range from 15 Hz.
Error
Query
Response
:SYSTem:FREQuency?
:SYSTEM:FREQUENCY 15HZ
(Headers : ON)
15HZ
(Headers : OFF)
The frequency range has been set to from 15 Hz.
An execution error occurs in the following cases :
• When measurement mode is selected (Command)
• When a mode other than "leakage current mode" is set (Command)
Note
Use this command after using the
:MODE
command to turn off (and no mea-
surement mode is selected).
Setting the frequency range to 0.1 Hz will activate the following settings.
• Leakage current mode: When measured current is ACPeak, ACDC is used.
In the AUTO range, HOLD3 is used.
• Voltmeter mode: In AUTO range, HOLD3 is used.
• Protective conductor ammeter mode: Frequency range setting is disabled
and operation starts from a 15 Hz setting.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......