8.7 Command Message Description
216
Error
If the set data unit number is greater than the total number of data units, an exe-
cution error occurs.
Queries the equipment name/control number of the saved data
Reads saved data
Syntax
Query
:MEMory:READ:MEASure?
<numerical value>,<characters>
<numerical value>
= Data unit number (NR1)
: Numeric value from 1 to the number of models (total number of
data units)
<characters>
= Measurement mode
EARTh/ENCLosure1(ENCL1)/ ENCLosure2(ENCL2)/
ENCLosure3(ENCL3)/PATient1(PAT1)/
PATient2(PAT2)/PATient3(PAT3)/PAUXiliary/
TOUCh1(TOUC1)/TOUCh2(TOUC2)/
TOUCh3(TOUC3)/PATientP2E(PATP2E)/
PATientSIPSOP(PATSIPSOP)/
PATientFTYPE(PATFTYPE)/PATientMP(PATMP)/
TPATientP2E(TPATP2E)/TPATientSIPSOP(TPATSIPSOP)/TPA-
TientFTYPE(TPATFTYPE)/TPATientMP(TPATMP)/FREE
When network A, D, E, or F is selected
EARTh
:Earth leakage current
ENCLosure1
:Enclosure - Earth leakage current
ENCLosure2
:Enclosure - Enclosure current
ENCLosure3
:Enclosure - Line leakage current
FREE
:Free current (only network A)
When network B1 is selected(only ST5540)
EARTh
:Earth leakage current
ENCLosure1
:Enclosure - Earth leakage current
ENCLosure2
:Enclosure - Enclosure current
PATient1
:Patient leakage current I
PATient2
:Patient leakage current II
PATient3
:Patient leakage current III
PAUXiliary
:Patient auxiliary current
FREE
:Free current
When network C or G is selected
EARTh
:Earth leakage current
TOUCh1
:Touch current Enclosure - Earth
TOUCh2
:Touch current Enclosure - Enclosure
TOUCh3
:Touch current Enclosure - Line
When network B2 is selected(only ST5540)
EARTh
:Earth leakage current
TOUCh1
:Touch current Enclosure - Earth
TOUCh2
:Touch current Enclosure - Enclosure
PATAUXiliary
:Patient auxiliary current
PATientP2E
:Patient leakage current Patient connection - Earth
PATientSIPSOP
:Patient leakage current SIP/SOP
PATientFTYPE
:Patient leakage current F-type applied part
PATientMP
:Patient leakage current Metal accessible part
TPATientP2E
:Total patient leakage current Applied part- Earth
TPATientSIPSOP
:Total patient leakage current SIP/SOP
TPATientFTYPE
:Total patient leakage current F-type applied part
TPATientMP
:Total patient leakage current Metal accessible
part
FREE
:Free current
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......