10.14 Measurement Networks
274
Network D is a measurement network that complies with UL.
(1)
Circuit Configuration
When network D is selected, the instrument can also be used in compliance with
the following standards on testing leakage current :
(2)
There are many other applicable standards.
Network D (for UL)
0.15
F
±
2%
1.5 k
±
1%
Network (1.5 k
) with 1.5 k
and 0.15
F
V
Applicable standards
Applicable UL standards
(e.g., UL 471, UL 1310, UL 1437, UL 1492)
Measurement circuit configuration
Network with 1.5 k
and 0.15
F
(when input protective fuse is short-circuited)
Element tolerance
Resistance: ± 1%
Capacitor : ± 2%