8.7 Command Message Description
202
Sets and queries the measurement terminal of the Patient leakage current
(Patient
–
Earth) and the total patient leakage current (Patient
–
Earth)
(only ST5540)
Syntax
Command
Query
Response
:CONFigure:TERMinal
<characters>
:CONFigure:TERMinal?
<characters>
<characters>
= T1T2/T2
T1T2:Terminal T1 and Terminal T2
T2:Terminal T2 and grounding terminal
Function
Command
Sets the measurement terminal of the Patient leakage current (Patient - Earth)
and the total patient leakage current (Patient - Earth)
Query
Returns the setting of the measurement terminal of the Patient leakage current
(Patient - Earth) and the total patient leakage current (Patient - Earth) in charac-
ters.
Example
Command
:CONFigure:TERMinal T1T2
Sets the measurement terminal of the Patient leakage current (Patient - Earth)
and the total patient leakage current (Patient - Earth) to T1 and T2 terminals.
Query
Response
:CONFigure:TERMinal?
:CONFIGURE:TERMINAL T1T2
(Headers: ON)
T1T2
(Headers: OFF)
Measurement terminal of the Patient leakage current (Patient - Earth) and the
total patient leakage current (Patient - Earth) are set to T1 and T2 terminals.
Error
An execution error occurs in the following cases:
• When measurement mode is other than the patient leakage current (Patient -
Earth) or the total patient leakage current (Patient - Earth)
• When a mode other than "leakage ammeter mode" is set
• The grounding class of equipment under test is set to something other than
internally powered equipment
• During automatic measurements
Note
Use this command after executing the
:MODE
command to activate the mea-
surement mode.
Sets and queries the target voltage
Syntax
Command
Query
Response
:CONFigure:VOLTage
<characters>
:CONFigure:VOLTage?
<characters>
<characters>
= ACDC/ AC/ DC/ACPEAK
ACDC
: AC+DC(alternating current and direct current)
AC
: Alternating current
DC
: Direct current
ACPEAK
: AC peak
Function
Command
Sets the target voltage.
Query
Returns the target voltage setting as alphabetic characters.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......