8.7 Command Message Description
198
When network B2 is selected
Example
Command
:CONFigure:AUTO OFF;:CONFigure:SPECial NAPPly
Specific manual measurement: Sets the application of 110% of rated voltage as
positive phase.
Query
Response
:CONFigure:AUTO OFF;:CONFigure:SPECial?
:CONFIGURE:SPECIAL NAPPLY
(Headers : ON)
NAPPLY
(Headers : OFF)
Specific manual measurement: The application of 110% of rated voltage has
been set to positive phase.
Error
An execution error occurs in the following cases:
• When the measurement method is set to "automatic measurement"
• When a mode other than "leakage current mode" is set
• When a Network other than "B1" or "B2" is set
• When a measurement mode other than Patient leakage current, F-type
applied part, Patient leakage current : Metal accessible part, Total patient
leakage current : F-type applied part, Total patient leakage current : Metal
accessible part is set
Note
• Use this command after executing the
:MODE
command to activate the mea-
surement mode.
• Use this command after using the
:NETWork
command to enter the "B1" or
"B2" network.
• Use this command after executing the
:MODE
command to activatePatient
leakage current:F-type applied part, Patient leakage current : Metal accessi-
ble part, Total patient leakage current : F-type applied part, Total patient leak-
age current : Metal accessible part.
Measurement modes
Special test condition (Specific applied voltage)
Class-I equipment
Class-II equipment
Internally powered
equipment
Application
of 110%
voltage:
Positive
phase
Application
of 110%
voltage:
Negative
phase
Application
of 110%
voltage:
Positive
phase
Application
of 110%
voltage:
Negative
phase
Application
of 110%
voltage:Pos-
itive phase
Application
of 110%
voltage:
Negative
phase
Earth leakage current
–
–
–
–
–
–
Touch current :
Enclosure - Earth
–
–
–
–
–
–
Touch current :
Enclosure - Enclosure
–
–
–
–
–
–
Patient leakage
current :Patient - Earth
–
–
–
–
–
–
Patient leakage
current :SIP/SOP
–
–
–
–
–
–
Patient leakage
current :
F-type applied part
Patient leakage cur-
rent :
Metal accessible part
: Can be set, –: Cannot be set
Specific manual measurement: Sets and queries the application of 110% of
rated voltage(only ST5540)
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......