10.7 User interface
264
10.7 User interface
Display
320 x 240 dot matrix LCD (with backlight)
LCD contrast adjustment : Provided with front panel control
Operation panel
6 x 6 matrix touch panel
Key lock function provided (asserts KEYLOCK terminal on EXT I/O)
10.8 EXT I/O
External I/O (standard)
• Connector : 36-pin, female
• Isolation (output: in 3.3 V power supply, internal GND (equivalent
to chassis grounding)
• Negative logic, open collector output
• Input:START
Measurements start when this signal is set to LO
STOP
Terminated when set to LO
KEYLOCK Disable all except Start switch
LOAD0(to 4) Reads saved panel (30 panels)
(EXT.DCV) External power 5 to 24 V DC input
(EXT.COM)External COM input
• Output:TEST
Outputs LO during measurements
(automatic measurement only)
MEAS
When there are multiple measurement items,
an equivalent number of outputs are generated
PASS
Outputs PASS judgment result for each
measurement item
FAIL
Outputs FAIL judgment result for
each measurement item
LOW
Outputs LOW judgment result for each measurement
item.Continues output if a single LOW state is
produced during automatic tests produced during
automatic tests
T-FAIL
Generates output even if one FAIL judgment
result is produced during automatic tests
(INT.DCV)
Internal 5 V DC output
(isolated from internal circuits)
(INT.GND)
Internal GND output
(equivalent to chassis grounding)
10.9 Relay output for medical equipment
(only ST5540)
Relay output
S10 : Relay between functional earth terminal and measurement power supply system
ground
S12 : Relay between patient connection and measurement power supply circuit ground
S13 : Relay between metal accessible part not protectively earthed and earth
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......