8.7 Command Message Description
195
8
Chapter 8
Contro
lling the
Instru
ment From
a Comp
uter
Sets and queries the set measurement current
(protective conductor ammeter)
Syntax
Command
Query
Response
:CONFigure:PCC
<characters>
:CONFigure:PCC?
<characters>
<characters>
= ACDC/AC/DC/ACPEAK
ACDC : AC+DC(alternating current and direct current)
AC : Alternating current
DC : Direct current
ACPEAK : ACPeak
Function
Example
Command
Query
Sets protective conductor current.
Returns protective conductor current setting as alphabetic characters.
Command
:CONFigure:PCC ACDC
Sets protective conductor current to AC + DC.
Error
Query
Response
:CONFigure:PCC?
:CONFIGURE:PCC ACDC
(Headers : ON)
ACDC
(Headers : OFF)
Protective conductor current has been set to AC + DC.
Mode is set to something other than "protective conductor ammeter mode".
Setting ACPeak when the frequency range is set between 0.1 Hz and 1 MHz
causes an execution error.
Note
Use this command after using the
:SYSTem:MODE
command to set the pro-
tective conductor ammeter mode.
Setting a frequency range from 0.1 Hz will activate the following settings.
•
Leakage current mode: When measured current is ACPeak, ACDC is used.
•
In the AUTO range, HOLD3 is used.
•
Voltmeter mode: In AUTO range, HOLD3 is used.
•
Protective conductor ammeter mode: Frequency range setting is disabled
and operation starts from a 15 Hz setting.
Sets and queries current range (protective conductor ammeter)
Syntax
Command
Query
Response
:CONFigure:PCC:RANGe
<characters>
:CONFigure:PCC:RANGe?
<characters>
<characters>=
HOLD1/HOLD2
HOLD1:50mA range
HOLD2:10mA range
Function
Example
Command
Query
Sets the current range for protective conductor ammeter.
Returns the current range for protective conductor ammeter setting as alpha-
betic characters.
Command
:CONFigure:PCC:RANGe HOLD1
Sets the 50 mA range as current range for protective conductor ammeter.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......