4.2 Selecting the Grounding Class of Equipment Under Test
49
3
Cha
pter 4
Se
ttings
4
4.2 Selecting the Grounding Class of
Equipment Under Test
Initial screen
1.
On the initial screen, press the
key to display the setting screen of the equip-
ment under test.
2.
Set the grounding class of the equipment under test.
3.
To manage measurement data, register equipment name and control number.
See
"Registering an Equipment Name/Control Number" (p. 50)
4.
Press
to return to the initial screen.The keys displayed indicate inspection items required
by the selected class.
Setting screen of equipment under test
(when selecting network B1 or B2)
2
2
3
3
4
4
Inspection of equipment with Type
B applied part
(when selecting network B1, B2)
Inspection of equipment with Type
BF applied part
(when selecting network B1, B2)
Inspection of equipment with Type
CF applied part
(when selecting network B1, B2)
Inspection of Class-I equipment
Inspection of Class-II equipment
Inspection of internally powered
equipment
Setting screen of equipment under test
(when selecting network A, C, D, E, F or G)
1
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......