8.7 Command Message Description
203
8
Chapter 8
Contro
lling the
Instru
ment From
a Comp
uter
Example
Command
:CONFigure:VOLTage ACDC
Sets the target voltage to AC+DC.
Query
Response
:CONFigure:VOLTage?
:CONFIGURE:VOLTAGE ACDC
(Headers : ON)
ACDC
(Headers : OFF)
The target voltage has been set to AC+DC.
Error
An execution error occurs when mode is set to something other than "voltmeter
mode."
Note
Use this command after executing the
:SYSTem:MODE
command to set the
instrument into voltmeter mode.
Sets and queries the target voltage
Sets and queries the voltage range
Syntax
Command
Query
Response
:CONFigure:VOLTage:RANGe
<characters>
:CONFigure:VOLTage:RANGe?
<characters>
<characters>
= AUTO/ HOLD1/ HOLD2/ HOLD3/ HOLD4
AUTO : Auto range
HOLD1 : 50.00 V range
HOLD2 : 5.000 V range
HOLD3 : 500.0 mV range
HOLD4 : 50.00 mV range
Function
Command
Sets the voltage range.
Query
Returns the voltage range setting as alphabetic characters.
Example
Command
:CONFigure:VOLTage:RANGe AUTO
Sets the voltage range to the Auto range.
Query
Response
:CONFigure:VOLTage:RANGe?
:CONFIGURE:VOLTAGE:RANGE AUTO
(Headers: ON)
AUTO
(Headers: OFF)
The voltage range has been set to the Auto range.
Error
An execution error occurs when mode is set to something other than "voltmeter
mode."
Note
Use this command after executing the
:SYSTem:MODE
command to set the
instrument into voltmeter mode.
Setting auto range when the frequency range is set between 0.1 Hz and 1 MHz
causes an execution error.
Sets and queries the wait time for switching operation in automatic measure-
ment (delay c)
Syntax
Command
Query
Response
:CONFigure:WTIMe:ETC
<numerical value>
:CONFigure:WTIMe:ETC?
<numerical value>
<numerical value>
= 1 to 1800 (NR1)
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......