9.5 Timing Chart
251
8
9
Chapte
r 9 Ext
ernal
Co
ntrol
This is a timing chart for output signals during manual measurements.
9.5 Timing Chart
Description
Time
MIN
MAX
T1
Pulse width of measured signal
55ms
–
T2
Judgment result output time, display update
time
Other than
ACpeak:550ms
–
ACpeak:600ms
–
•
Free run measurements are made while in the leakage current measurement
screen.
•
Calibration is performed once a minute. Measurements are temporarily halted
during calibration. (The MEAS signal goes HI)