10.6 System-related Functions
263
8
9
Chap
ter 10
S
pe
c
ificat
ions
10
10.6 System-related Functions
Beep sound setting
(a) Allowable value judgment :
Selectable from "ON when Pass," "ON when FAIL," and "OFF"
(b) Key input : Selectable from ON/OFF
(c) T3 (110% voltage application terminal) output : Selectable from ON/OFF
*
(d) Line voltage output from T2 : Selectable from ON/OFF
Save/load function
30 panels for saving the following setting data
(measurement mode, network, equipment name, control number, grounding
class, applied part, measurement range, filter, target current, allowable value
setting, fault condition setting, power supply polarity switching, automatic
measurement items, automatic measuring time, measurement delay time,
Panel name)
Data save function
Saved content
: Equipment under test information (equipment name,
serial number), measurement data, date
Memory capacity : Data on up to 100 units
Clock function
Auto calendar, automatic leap-year adjustment, 24 hour clock
Clock accuracy: Deviation of about 4 minutes a month
Data backup function
SRAM (Set condition), RTC
Backup battery life: Approx. 10 years (reference value at 25
°C
(77
°F
))
Backlight automatic OFF
function
(a) Constant ON (default)
(b) Auto OFF (1 min. to 30 min., in 1 min. increments)
The backlight automatically turns off after the set time elapses, turns ON
when a key is pressed, then turns OFF again after the set time elapses.
Self-test function
(a) MEM (internal RAM)
(b) KEY (6 x 6 matrix touch panel)
(c) LCD (front LCD panel)
(d) LED (warning lamp, LCD backlight)
(e) Beep sound
Language setting
(a) Japanese
(b) English (default)
Fuse blown check function
Checks for blown fuses in the network
At power on (when set) or key operation in the system screen
Relay check function
Checks whether the network relay works normally
At power on (when set) or key operation in the system screen
Allowable lower limit value
setting
Sets a lower limit value for all measurements.
System reset
(a) Returns instrument to its factory defaults
(includes measurement data, setting conditions, etc.)
(b) Clears all measurement data.
(c) Clears all condition setting data including panels.
*
only ST5540
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......