8.7 Command Message Description
236
Function
Example
Command
Query
Sets the allowable value (lower limit) to On/Off.
Returns all allowable value (lower limit) On/Off settings as alphabetic charac-
ters.
Command
:SYSTem:LOWer ON
Sets all allowable values (lower limit) On/Off to On.
Error
Query
Response
:SYSTem:LOWer?
:SYSTEM:LOWER ON
(Headers: ON)
ON
(Headers: OFF)
All allowable values (lower limit) On/Off has been set to On.
An execution error occurs in the following cases:
• When measurement mode is selected (Command)
• When a mode other than "leakage current mode" is set (Command)
Note
Use this command after using the
:MODE
command to turn off (and no mea-
surement mode is selected).
Turning On/Off an allowable value (lower limit) after all allowable values (lower
limit) have been turned On/Off will not result in a query asking to turn On/Off all
allowable values.Use the query command for turning an allowable value On or
Off to check.
Sets and queries all allowable values (lower limit) On/Off
Sets and queries the coefficients for all allowable values (lower limit)
Syntax
Command
Query
Response
:SYSTem:LOWer:COEFficient
<numerical value>
:SYSTem:LOWer:COEFficient?
<numerical value>
<numerical value>
=5 to 99(NR1)
Function
Example
Command
Query
Sets the coefficient for all allowable values (lower limit).
Returns the set coefficients for all allowable values (lower limit) as a numeric
value.
Command
:SYSTem:LOWer:COEFficient 10
Sets coefficients for all allowable values (lower limit) to 10%.
Error
Query
Response
:SYSTem:LOWer:COEFficient?
:SYSTEM:LOWER:COEFFICIENT 10
(Headers: ON)
10
(Headers : OFF)
Coefficient settings for all allowable values (lower limit) are set to 10%.
An execution error occurs in the following cases :
• When measurement mode is selected (Command)
• When a mode other than "leakage current mode" is set(Command)
Note
• Use this command after using the
:MODE
command to turn off (and no mea-
surement mode is selected).
• The unit is [%]. Fractions are rounded up when 0.5 or greater and rounded off
when less than 0.5.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......