
1.6 Types of Leakage Current Measurement
16
This is electric current that flows through an "
applied part
human body
ground" path.
Measure medical instruments with a
signal input/output section
and
applied parts
other than F-type. Mea-
sure assuming that medical equipment with a broken
signal input/output section
is connected. Instead of
connecting damaged medical equipment to generate a
single-fault condition
, input a voltage that is 110%
of the rated voltage.
(Use the T3 terminal on the instrument)
IEC 60601-1:2005 3rd Edition stipulates that the connection of damaged medical equipment does not
constitute a
single-fault condition
, but handles it as a
normal condition
.
Patient leakage current measurement (external voltage on a SIP/SOP)
JIS T0601-1:1999 treats patient leakage current II as patient leakage current I plus a
single fault condition
.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......