6.6 Printing Measurement Data (as required)
108
The maximum value, measurement conditions, and other data can be printed when the optional
9442 Printer is connected with the 9444 Connection cable.
Necessary items
• 9442 Printer
• 9443-01 AC Adapter with power cord (for Japan)
• 9443-02 AC Adapter (for EU)
• 1196 Recording paper
To connect the main instrument and printer :
• 9444 Connection Cable
Setting of the
9442 Printer
Change the settings of the software dip switches (DIP SW) to use the 9442 for the
ST5540 or ST5541.
• The 9442 is shipped with the function settings for use with the Hioki 3166 Clamp on power HiT-
ESTER. Before using, always change the settings of the DIP switches.
• For details on the operations and handling of the printer, refer to the operation manual supplied
to the printer.
• For the printer, use the 1196 Recording paper (thermal paper, 10 rolls) or an equivalent.
Procedure
1.
Turn off the power of the 9442.
2.
Turn on the power while pressing the ON LINE button. Release the button after a list of the cur-
rent settings starts printing out.
The print out of the current settings is followed by the prompt:
Continue? :Push 'On-line SW'
Write?
:Push 'Paper feed SW'
3.
Press the ON LINE button to change the settings.
"Dip SW-1" is printed to make a settings for DIP SW 1.
6.6 Printing Measurement Data (as required)
Setting up and connecting a printer
Preparation before Connection
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......